User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: Granville-Phillips 270:Gauge Controller DATE: 17-Nov-95 AUTHOR: User Contributed REVISION: ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 12 NUMBER OF LINES: 103 CONFIGURATION: Fluke 5700A STANDARD: ESI SR1 1Mê(1) ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- R N P U M F W 1.002 ASK+ X 1.003 STD ESI SR1 1Mê(1) 1.004 HEAD ÍÍ INITIAL CONDITIONS ÍÍ 1.005 DISP With UUT off, mechanically zero all the meters. 1.006 DISP Connect the UUT power cord to an isolation transformer. 1.007 DISP Connect the 5700a V/ê/A HI output, through the 1Mê 1.007 DISP standard resistor, to the HI of the COLLECTOR BNC 1.007 DISP on the UUT. 1.007 DISP 1.007 DISP Connect the LO side of the UUT BNC to the LO OUTPUT 1.007 DISP of the 5700A. 1.008 DISP On the UUT set the FILAMENT and EMISSION swithes down. 1.009 DISP Turn the UUT and allow to warm up for fifteen minutes. 1.010 HEAD {} 1.011 HEAD {ÉÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍ»} 1.012 HEAD {º Pressure Meter Verification º} 1.013 HEAD {ÈÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍͼ} 1.014 HEAD ÍÍ Pressure Meter Verification ÍÍ 1.015 JMP 2.001 1.016 EVAL 2.001 DISP Set the UUT AUTORANGE off, and the RANGE to 10E-8 TORR. 2.002 MESS Adjust the 5700 for a reading of "12" on the UUT. 2.003 5700 2.4000mV N 2W 2.004 MEME 2.005 MATH mem=2.4 2.006 MEMC nA 0.120U ÿ10e-8 3.001 MESS Adjust the 5700 for a reading of "6" on the UUT. 3.002 5700 1.2000mV N 2W 3.003 MEME 3.004 MATH mem=1.2 3.005 MEMC nA 0.120U ÿ10e-8 4.001 HEAD {} 4.002 HEAD ÍÍ Pressure Meter Verification ÍÍ 4.003 DISP Set the UUT RANGE switch to 10E-7 TORR. 4.004 MESS Adjust the 5700 for a reading of "12" on the UUT. 4.005 5700 24.000mV N 2W 4.006 MEME 4.007 MATH mem=24 4.008 MEMC nA 1.20U ÿ10e-7 5.001 MESS Adjust the 5700 for a reading of "6" on the UUT. 5.002 5700 12.000mV N 2W 5.003 MEME 5.004 MATH mem=12 5.005 MEMC nA 1.20U ÿ10e-7 6.001 HEAD {} 6.002 HEAD ÍÍ Pressure Meter Verification ÍÍ 6.003 DISP Set the UUT RANGE switch to 10E-6 TORR. 6.004 MESS Adjust the 5700 for a reading of "12" on the UUT. 6.005 5700 240.00mV N 2W 6.006 MEME 6.007 MATH mem=240 6.008 MEMC nA 12.0U ÿ10e-6 7.001 MESS Adjust the 5700 for a reading of "6" on the UUT. 7.002 5700 120.00mV N 2W 7.003 MEME 7.004 MATH mem=120 7.005 MEMC nA 12.0U ÿ10e-6 8.001 HEAD {} 8.002 HEAD ÍÍ Pressure Meter Verification ÍÍ 8.003 DISP Set the UUT RANGE switch to 10E-5 TORR. 8.004 MESS Adjust the 5700 for a reading of "12" on the UUT. 8.005 5700 2.400V N 2W 8.006 MEME 8.007 MATH mem=2.4 8.008 MEMC uA 0.120U ÿ10e-5 9.001 MESS Adjust the 5700 for a reading of "6" on the UUT. 9.002 5700 1.200V N 2W 9.003 MEME 9.004 MATH mem=1.2 9.005 MEMC uA 0.120U ÿ10e-5 10.001 HEAD {} 10.002 HEAD ÍÍ Pressure Meter Verification ÍÍ 10.003 DISP Set the UUT RANGE switch to 10E-4 TORR. 10.004 MESS Adjust the 5700 for a reading of "12" on the UUT. 10.005 5700 24.00V N 2W 10.006 MEME 10.007 MATH mem=24 10.008 MEMC uA 3.60U ÿ10e-4 11.001 MESS Adjust the 5700 for a reading of "6" on the UUT. 11.002 5700 12.00V N 2W 11.003 MEME 11.004 MATH mem=12 11.005 MEMC uA 3.60U ÿ10e-4 12.001 DISP Disconnect all equipment and turn off UUT. 12.002 HEAD {} 12.003 DISP {ÉÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍ»} 12.003 DISP {º THIS COMPLETES THE VERIFICATION º} 12.003 DISP {ÈÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍͼ} 12.004 END